Dimension Vista - Drugs of Abuse Controls (2017-12-21)

Starting date:
December 21, 2017
Posting date:
January 12, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65682

Dimension Vista - Drugs of Abuse Controls

Siemens Healthcare Diagnostics has determined the Phencyclidine (PCP) analyte in Drugs of Abuse Positive Control [UDAT CON(+)] Lot 7cqp15 is recovering lower than expected. Customers may occasionally get a negative result for PCP with UDAT CON(+) Lot 7cqp15. All other analytes in the UDAT CON(+) recover appropriately.

 

7CQP15

KC515

Manufacturer

Siemens Healthcare Diagnostics Inc.
500 GBC DRIVE, MAILSTOP 514, P.O. BOX 6101
Newark
19714
Delaware
UNITED STATES